Sorting Based on Measured Result (Comparator Function)

Setting Up Comparator Function

Setting Up Comparator Function

Turning ON/OFF comparator function

To enable/disable the comparator function, use the following command.

Setting up limit range

Clearing (resetting) limit range
You can clear the ON/OFF state and the lower and upper limit values of all limit ranges
(BIN1 to BIN9 and the secondary parameter limit range) and the ON/OFF state of AUX
BIN to recover the factory-shipped default values (refer to Table D-1, "Initial settings,
settings that can be saved/recalled, settings that can be backed up," on page 259).
To clear the limit ranges, use the following command.
Selecting limit range designation method
You can select the designation method of the primary parameter limit ranges (BIN1 to
BIN9) from the following three modes:
NOTE
Only the absolute mode can be used as the designation method of the secondary parameter
limit range.
To select the designation method, use the following command.
To set up the reference value for the absolute tolerance mode or percent tolerance mode,
use the following command.
82
:CALC:COMP on page 134
:CALC:COMP:CLE on page 136
Mode
Absolute mode
Absolute tolerance mode
Percent tolerance mode
*1.Boundary value - reference value
*2.((Boundary value - reference value)/reference value)× 100
:CALC:COMP:MODE on page 141
:CALC:COMP:PRIM:NOM on page 144
Description
Designation using the absolute value
Designation using the
relative value (deviation
from the reference value)
Designation using the absolute
*1
value
Designation using the percentage of
*2
the reference value
Chapter 7