Agilent Technologies 4288A Programming Manual page 106

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Measurement Applications (Sample Programs)
Basic Capacitor Measurement
560
!=============================================
570
DEF FNCompen(@Agt4288a,Standard$,Limit)
580
DIM Inp_char$[9],Buff$[9],Std$[9],Err$[50]
590
REAL Curr_freq,Freq(1:2),Para1,Para2,Zm,Ym,Gm,Bm
600
REAL Cpref,Dref,Zref,Gref,Bref
610
INTEGER Err_flag
620
OUTPUT @Agt4288a;":SOUR:FREQ?"
630
ENTER @Agt4288a;Curr_freq
640
Freq(1)=1.E+3
650
Freq(2)=1.E+6
660
Load_para$="CPD"
670
OUTPUT @Agt4288a;":STAT:OPER:ENAB 128"
680
OUTPUT @Agt4288a;"*SRE 128"
690
SELECT Standard$
700
CASE "Open"
710
Std$="STAN1"
720
CASE "Short"
730
Std$="STAN2"
740
CASE "Load"
750
Std$="STAN3"
760
CALL Inp_data("Load(Cp) Value @1kHz",Load1(1))
770
CALL Inp_data("Load(D)
780
CALL Inp_data("Load(Cp) Value @1MHz",Load1(2))
790
CALL Inp_data("Load(D)
800
OUTPUT @Agt4288a;":CORR:CKIT:STAN3:FORM "&Load_para$
810
FOR I=1 TO 2
820
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
830
OUTPUT @Agt4288a;":CORR:CKIT:STAN3 ";Load1(I);",";Load2(I)
840
NEXT I
850
END SELECT
860 Compen_meas:
!
870
PRINT "Set "&Standard$&"-Connection."
880
INPUT "OK? [Y/N]",Inp_char$
890
IF UPC$(Inp_char$)="Y" THEN
900
FOR I=1 TO 2
910
PRINT "Frequency:";Freq(I)
920
OUTPUT @Agt4288a;":SOUR:FREQ ";Freq(I)
930
OUTPUT @Agt4288a;"*CLS"
940
OUTPUT @Agt4288a;"*OPC?"
950
ENTER @Agt4288a;Buff$
960
! Measurement
970
ON INTR 7 GOTO Meas_end
980
ENABLE INTR 7;2
990
OUTPUT @Agt4288a;":CORR:COLL "&Std$
1000 Meas_wait: GOTO Meas_wait
1010 Meas_end: OFF INTR 7
1020
! Error Check
1030
OUTPUT @Agt4288a;":SYST:ERR?"
1040
ENTER @Agt4288a;Err_no,Err$
1050
IF Err_no<>0 THEN
1060
PRINT "Error: "&Err$
1070
GOTO Compen_meas
1080
END IF
1090
! Data Check
1100
OUTPUT @Agt4288a;":CORR:DATA? "&Std$
1110
ENTER @Agt4288a;Para1,Para2
1120
Err_flag=0
1130
SELECT Standard$
1140
CASE "Open"
1150
Ym=SQRT(Para1*Para1+Para2*Para2)
1160
PRINT "G =";Para1,"B =";Para2,"|Y| =";Ym
1170
IF Ym>=Limit THEN Err_flag=1
1180
CASE "Short"
1190
Zm=SQRT(Para1*Para1+Para2*Para2)
104
Value @1kHz",Load2(1))
Value @1MHz",Load2(2))
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents