Turning ON/OFF limit range
To turn ON/OFF the primary parameter limit range (BIN1 to BIN9), use the following
command.
:CALC:COMP:PRIM:BIN{1-9}:STAT on page 143
Only BINs set to ON with the above command become the targets of sorting judgment. In
other words, even if the measured result falls within the limit range of a BIN set to OFF, it
will not be sorted to that BIN.
To turn ON/OFF the secondary parameter limit range, use the following command.
:CALC:COMP:SEC:STAT on page 146
If you specify OFF with the above command, no sorting judgment is made for the
measured result of the secondary parameter. The comparator sorting judgment result is
determined only on the basis of the measured primary parameter result.
Setting the range's lower and upper limit values
To set the lower limit value and upper limit value of the primary parameter limit range
(BIN1 to BIN9), use the following command.
:CALC:COMP:PRIM:BIN{1-9} on page 142
To set the lower limit value and upper limit value of the secondary parameter limit range,
use the following command.
:CALC:COMP:SEC:LIM on page 145
NOTE
If you set the upper limit value to a value equal to or less than the lower limit value, no
limit range is used. Doing this would be equivalent to setting the limit range to OFF.
As shown in Figure 7-1, "Sorting judgment flow," on page 87, the sorting judgment is
performed in increasing order of BIN number. Therefore, when the limit ranges of two
BINs overlap, the result is sorted to the BIN with the smaller number.
For the tolerance mode, the reference value does not need to be within the limit range
(between the lower limit value and upper limit value).
Gaps are allowed between the limit ranges.
Chapter 7
Sorting Based on Measured Result (Comparator Function)
Setting Up Comparator Function
83