:TRIG

Syntax
:TRIGger[:SEQuence1][:IMMediate]
Description
Immediately generates a trigger and executes a measurement regardless of the setup of the
trigger mode.
If the trigger system is not in the trigger wait state (trigger event detection state), executing
this command causes an error and the command is ignored.
For details on the trigger system, refer to "Trigger system" on page 52. (No query)
Equivalent key
No equivalent key is available on the front panel.
sequence

:TRIG:DEL

Syntax
:TRIGger[:SEQuence1]:DELay <numeric>[MS|M|S]
:TRIGger[:SEQuence1]:DELay?
Description
Specifies the waiting time between when a trigger is inputted and when the measurement
starts (trigger delay time).
Parameters
Description
Range
Initial value
Unit
Resolution
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that a fraction below the resolution is rounded off.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Equivalent key
[Delay]
sequence
Chapter 10
Trigger delay time
0 to 1
0
s (second)
1E-3
(
-
)
[Blue]
[Trig Mode]
Command Reference
<numeric>