Command Reference

:CALC2:FORM

:CALC2:FORM
Syntax
:CALCulate2:FORMat {D|Q|G|RP|RS}
:CALCulate2:FORMat?
Description
Specifies the secondary parameter to be measured.
If the primary parameter has been set to CP and the secondary parameter is set to RS, the
primary parameter is automatically changed to CS. Also, if the primary parameter has been
set to CS and the secondary parameter is set to G or RP, the primary parameter is
automatically set to CP.
Parameters
Query response
{D|Q|G|RP|RS}<newline><^END>
Related commands :CALC1:FORM on page 147
Equivalent key
[Meas Prmtr]
sequence

:CALC2:MATH:EXPR:CAT?

Syntax
:CALCulate2:MATH:EXPRession:CATalog?
Description
Reads out available parameters used when executing the command to specify the
expression of the secondary parameter in the deviation measurement mode (the
:CALC1:MATH:EXPR:NAME command). The query response is always DEV,PCNT.
(Query only)
Query response
DEV,PCNT<newline><^END> (fixed)
Related commands :CALC2:MATH:EXPR:NAME on page 151
Equivalent key
No equivalent key is available on the front panel.
sequence
150
D (initial value)
Specifies the dissipation factor as the secondary parameter.
Q
Specifies the quality factor (inverse value of D) as the secondary
parameter.
G
Specifies the equivalent parallel conductance measured using the
parallel equivalent circuit model as the secondary parameter.
RP
Specifies the equivalent parallel resistance measured using the
parallel equivalent circuit model as the secondary parameter.
RS
Specifies the equivalent series resistance measured using the series
equivalent circuit model as the secondary parameter.