Init; Init:cont - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

:INIT

Syntax
:INITiate[:IMMediate]
Description
Starts the trigger system once.
If the trigger system is in the idle state, executing this command starts the trigger system
and, after measurement is executed once, returns the instrument to the idle state.
If the trigger system is not in the idle state or successive startup of the trigger system is
enabled (set to ON with the :INIT:CONT command), executing this command causes an
error and the command is ignored.
For details on the trigger system, refer to "Trigger system" on page 52. (No query)
Related commands :INIT:CONT on page 183
Equivalent key
No equivalent key is available on the front panel.
sequence

:INIT:CONT

Syntax
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description
Determines whether to start up the trigger system successively.
For details on the trigger system, refer to "Trigger system" on page 52.
Parameters
ON or 1 (initial value)
OFF or 0
This setup is initialized to ON when using the :SYST:PRES command and to OFF when
using the *RST command.
Query response
{1|0}<newline><^END>
Related commands *RST on page 126
Equivalent key
No equivalent key is available on the front panel.
sequence
Chapter 10
Description
Enables successive startup.
Disables successive startup.
Command Reference
:INIT
183

Hide quick links:

Advertisement

Table of Contents
loading