:CALC3:MATH:STAT

Syntax
:CALCulate3:MATH:STATe {ON|OFF|1|0}
:CALCulate3:MATH:STATe?
Description
Determines whether to use the monitor function of the current flowing through the DUT
during measurement.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands :CALC4:MATH:STAT on page 153
Equivalent key
[Level Mon]
sequence

:CALC4:MATH:STAT

Syntax
:CALCulate4:MATH:STATe {ON|OFF|1|0}
:CALCulate4:MATH:STATe?
Description
Determines whether to use the monitor function of the voltage applied to the DUT during
measurement.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Related commands :CALC3:MATH:STAT on page 153
Equivalent key
[Level Mon]
sequence
Chapter 10
Enables the current monitor function.
Disables the current monitor function.
(
-
)
[Blue]
[Show Setting]
Enables the voltage monitor function.
Disables the voltage monitor function.
(
-
)
[Blue]
[Show Setting]
Command Reference