Reading out measurement signal level monitor result

Reading out measurement signal level monitor result
To read out measurement signal level monitor result, use the following command:
:DATA? {IMON|VMON} on page 171
Example 6-8 shows a sample program to read out the measurement signal level monitor
result. This program is stored on the sample program disk under the filename
"lvl_mon.bas."
This program turns ON the measurement signal level monitor function, performs
measurement once, and reads out and displays the measurement signal level monitor result.
The program is detailed below.
Line 40
Sets the GPIB address.
Line 50
Sets the data transfer format to the ASCII format.
Lines 70 to 80
Turns ON the measurement signal level current & voltage monitor
function.
Line 90
Makes the setup display the measurement signal level monitor result in
the instrument setup display area.
Lines 110 to 120
Turns ON the trigger system continuous startup and sets the trigger
mode to GPIB trigger.
Lines 160 to 200
Makes the setup generate an SRQ when measurement finishes and
clears the status byte register and operation status event register.
Lines 240 to 250
Sets the branch destination of the SRQ interrupt and enables the SRQ
interrupt.
Lines 260 to 270
Triggers the measurement through GPIB and waits for the completion
of measurement.
Lines 290 to 300
Reads out the voltage monitor value of the measurement signal level.
Lines 310 to 320
Reads out the current monitor value of the measurement signal level.
Lines 360 to 410
Converts the voltage monitor value and current monitor value of the
measurement signal level into mV and µ A, respectively, and displays
them.
Chapter 6
Reading Out Measured Result
79