Reading Out Measured Result
Reading Out Measured Result
The program of Example 6-6 is detailed below.
Lines 50 to 60
Sets the GPIB address.
Line 70
Sets the data transfer format to the binary format.
Lines 90 to 100
Turns OFF the trigger system continuous startup and sets the trigger
mode to external trigger.
Lines 120 to 130
Reads out the ON/OFF state of the comparator function and sets it to
the Comp_flag variable.
Executes the :READ? command, displays the message to prompt the
Lines 170 to 180
user to input an external trigger, and waits until the external trigger is
inputted and the measurement finishes.
Lines 190 to 220
Reads out the header part.
Line 240
If the Comp_flag value is 1 (the comparator function is ON), reads out
the measurement status, measured primary parameter result, measured
secondary parameter result, and comparator sorting result.
Line 260
If the Comp_flag value is not 1 (the comparator function is OFF),
reads out the measurement status, measured primary parameter result,
and measured secondary parameter result.
Line 280
Reads out the message terminator at the end of the data.
Lines 320 to 340
Displays the measurement status, measured primary parameter result,
and measured secondary parameter result.
Line 360
If the Comp_flag value is 1 (the comparator function is ON), displays
the comparator sorting result.
Line 380
If the Comp_flag value is not 1 (the comparator function is OFF),
displays "Comparator : OFF."