Error Messages
Error number: -220
-220
Parameter error
An error not included in the error numbers between -221 and -229 has occurred during the
analysis of a program data element. This error occurs, for example, when you attempt to
specify invalid values (values not finite when converted to an R-X format impedance
value) as the LOAD correction data or LOAD correction reference data. If this error
occurs, the command is ignored. This error also occurs when you attempt to specify an
invalid LOAD correction reference value by using the front panel keys.
-221
Setting conflict
A program data element complying with the syntax standard has been analyzed, but the
4288A cannot execute it at present.
-222
Data out of range
A data element (that does not violate the standard) has been received out of the range
defined for the 4288A.
-223
Too much data
The received block, equation, or string type program data complies with the standard, but
the amount of data exceeds the limit that the 4288A can handle due to memory or
device-specific conditions related to memory.
-230
Data corrupt or stale
The data is invalid or a newly initiated read operation has not been completed since the
latest access.
-241
Hardware missing
The received command or Query complied with the standard but could not be executed due
to hardware-related reasons (for example, an option was not installed).
270
Parameter Type Setting
Primary
Secondary
Parameter
Parameter
Cp
Cp
Cp
Cp
Cs
Cs
Cs
Values not finite when converted to an R-X
format impedance value
D
Independent of D value, Cp is 0
Independent of Q value, Cp is 0, also
Q
Independent of Cp value, Q is 0
G
Cp and G are both 0
Rp
Independent of Cp value, Rp is 0
D
Independent of D value, Cs is 0
Independent of Q value, Cs is 0, also
Q
Q is 0 independent of Cs value
Rs
Cs is 0 independent of Rs value
Appendix F