:CORR:CKIT:STAN3

Syntax
[:SENSe]:CORRection:CKIT:STANdard3 <numeric 1>,<numeric 2>
[:SENSe]:CORRection:CKIT:STANdard3?
Description
Defines the values of the LOAD compensation standard for the parameters you specify by
using the :CORR:CKIT:STAN3:FORM command.
These are set as the standard values for the measurement frequency when executing the
command (set with the :SOUR:FREQ command).
When using the multi-compensation function (set to ON with the :CORR:MULT
command) with the channel-by-channel definition of the standard values enabled (set to
ON with the :CORR:MULT:CKIT:STAN3 command), the standard values for the selected
channel (selected with the :CORR:MULT:CHAN command) are set at execution of the
command.
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands :CORR:LOAD on page 160
:CORR:CKIT:STAN3:FORM on page 156
:CORR:MULT on page 161
:CORR:MULT:CHAN on page 161
:CORR:MULT:CKIT:STAN3 on page 162
Equivalent key
[Load]
sequence
Chapter 10
<numeric 1>
Value of the primary
parameter.
-999.999 to 999.999
100E-9
F (farad)
(
[Blue]
-
[6]
) -
CorVal
-
RefEnt
Command Reference
:CORR:CKIT:STAN3
<numeric 2>
Value of the secondary
parameter.
-99.9999E9 to 99.9999E9
0
Depends on the type of
parameter.
155