:CALC:COMP:PRIM:NOM

:CALC:COMP:PRIM:NOM
Syntax
:CALCulate:COMParator:PRIMary:NOMinal <numeric>
:CALCulate:COMParator:PRIMary:NOMinal?
Description
Specifies the reference value used when specifying the primary parameter limit range for
the comparator function. This value can be used when the limit range designation method
is set to either absolute tolerance mode or percent tolerance mode.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands :CALC:COMP:MODE on page 141
:CALC:COMP:PRIM:BIN{1-9} on page 142
Equivalent key
[Comprtr]
sequence
144
Description
The reference value used when specifying the primary parameter
limit range.
Range
-999.999 to 999.999
Initial value
0
Unit
F (farad)
(
-
) -
[Blue]
[1]
<numeric>
∆ ABS
∆ %
-
/
Limit
Chapter 10