Agilent Technologies 4288A Programming Manual page 107

1khz/1mhz capacitance meter
Table of Contents

Advertisement

1200
PRINT "R =";Para1,"X =";Para2,"|Z| =";Zm
1210
IF Zm>=Limit THEN Err_flag=1
1220
CASE "Load"
1230
OUTPUT @Agt4288a;":CORR:CKIT:STAN3?"
1240
ENTER @Agt4288a;Cpref,Dref
1250
Bref=2*PI*Freq(I)*Cpref
1260
Gref=Bref*Dref
1270
Zref=1/SQRT(Gref*Gref+Bref*Bref)
1280
Bm=2*PI*Freq(I)*Para1
1290
Gm=Bm*Para2
1300
Zm=1/SQRT(Gm*Gm+Bm*Bm)
1310
PRINT "Cpref=";Cpref,"Dref=";Dref,"|Zref| =";Zref
1320
PRINT "Cp
1330
IF ABS((Zm-Zref)/Zref)>=Limit THEN Err_flag=1
1340
END SELECT
1350
IF Err_flag<>0 THEN
1360
PRINT "Out of limit!!"
1370
GOTO Compen_meas
1380
END IF
1390
NEXT I
1400
PRINT Standard$&" Data Measurement Complete"
1410
OUTPUT @Agt4288a;":SOUR:FREQ ";Curr_freq
1420
RETURN 0
1430
ELSE
1440
PRINT "Program Interruption"
1450
OUTPUT @Agt4288a;":SOUR:FREQ ";Curr_freq
1460
RETURN -1
1470
END IF
1480 FNEND
1490
!=============================================
1500
! Data Input Function
1510
!=============================================
1520 SUB Inp_data(Mes$,Inp_val)
1530
DIM Inp_char$[30]
1540
ON ERROR GOTO Inp_start
1550 Inp_start:!
1560
PRINT "Input "&Mes$
1570
INPUT "Value?",Inp_char$
1580
Inp_val=VAL(UPC$(Inp_char$))
1590
PRINT "Input Value: ";Inp_val
1600
INPUT "OK? [Y/N]",Inp_char$
1610
IF UPC$(Inp_char$)<>"Y" THEN Inp_start
1620
OFF ERROR
1630 SUBEND
Chapter 9
Measurement Applications (Sample Programs)
Basic Capacitor Measurement
=";Para1,"D
=";Para2,"|Z|
=";Zm
105

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents