Comparator - Agilent Technologies 4288A Programming Manual

1khz/1mhz capacitance meter
Table of Contents

Advertisement

Functional Comparison and GPIB Command Correspondence

Comparator

Functional comparison
Sorting target parameter
Reverse parameter function
Using a deviation from
the reference value for
Limit mode (way of
designation
designating the limit
range)
Using an absolute
value for designation
Primary parameter
BIN sorting
Secondary parameter
AUX BIN function
BIN count function
Low C reject function
Differences in usage of GPIB commands
Sorting target
Can be set up independently of the
parameter
measurement parameter.
You cannot turn ON/OFF each BIN
separately. Sorting targets are all of the
BINs for which the limit range has been set
ON/OFF of BIN
up. Therefore, to exclude BINs that have
been set up from sorting targets, you need
to clear all the limit ranges and set them
again.
You use a single command to set up the
limit ranges for all BINs (sequential mode).
Setting limit ranges
The limit ranges of each BIN must be
in absolute values
continuous.
Executing the limit value setup command
(the boundary value to detect Low C) turns
ON/OFF of Low C
ON the Low C reject function.
reject function
Sorts into NPP reserved only for Low C.
Sorting result when
Low C occurs
UNVAL: Counts into OUT_OF_BINS.
BIN count function
Low C: No count
Appendix B
Information for Replacing 4278A with 4288A
4278A
Cp-D, Cp-Q, Cp-G, Cs-D,
Cs-Q, Cs-Rs
Available
% tolerance mode, absolute
tolerance mode
Sequential mode
9 BINs + OUT_OF_BINS
PASS/FAIL judgment
Available
Counting all sorting results,
counting for each channel
Select the detection limit value
from 1%, 2%, 3%, 4%, 5%, or
6%.
4278A
Depends on the measurement parameter
and cannot be set up independently.
You can turn ON/OFF each BIN separately.
Sorting targets are the BINs set to ON only.
Therefore, when setting up BINs, you need
to set up the limit range for each BIN and
then turn it ON.
You use several commands to set up the
limit range for each BIN separately
(absolute mode). The limit range of each
BIN does not have to be continuous (that is,
gaps are allowed).
Executing the limit value setup command
does not turn ON the Low C reject function
automatically. Therefore, you need to set
up the limit value and then turn ON the
Low C reject function.
Sorts into suitable BIN according to
measurement results. (The 4288A does not
have a BIN reserved for only Low C.)
OVLD: Counts into OVLD.
Low C: Counts into a suitable BIN
according to measurement results.
4288A
Cp-D, Cp-Q, Cp-G, Cp-Rp,
Cs-D, Cs-Q, Cs-Rs
N/A
% tolerance mode, absolute
tolerance mode
Absolute mode
9 BINs + OUT_OF_BINS
PASS/FAIL judgment
Available
Counting all sorting results,
counting for each channel
Set up the detection limit value
freely within the range of 0 to
10%.
4288A
221