Command Reference

:CALC:COMP:COUN:MULT:OVLD?

:CALC:COMP:COUN:MULT:OVLD?
Syntax
:CALCulate:COMParator:COUNt:MULTi:OVLD?
Description
Reads out the overload count value of the selected channel.
When the multi-compensation function is tuned off, reads out normal count value
(common value for all channels). In other words, this command has the same function as
the :CALC:COMP:COUN:OVLD? command. (Query only)
Query response
{numeric}<newline><^END>
Related commands :CALC:COMP:COUN:MULT:DATA? on page 139
:CORR:MULT on page 161
:CALC:COMP:COUN on page 137
:CALC:COMP:COUN:CLE on page 137
:CALC:COMP on page 134
Equivalent key
No equivalent key is available on the front panel.
sequence

:CALC:COMP:COUN:OVLD?

Syntax
:CALCulate:COMParator:COUNt:OVLD?
Description
Reads out each overload count value. (Query only)
Query response
{numeric}<newline><^END>
Related commands :CALC:COMP:COUN:DATA? on page 138
:CALC:COMP:COUN on page 137
:CALC:COMP:COUN:CLE on page 137
:CALC:COMP on page 134
Equivalent key
No equivalent key is available on the front panel.
sequence
140