Chapter 8: Daily Shutdown; Overview; Preparing The Cleaning Solution; Shutting Down The Instrument - Beckman Coulter CytoFLEX Series Instructions For Use Manual

Flow cytometers
Table of Contents

Advertisement

Overview

This chapter provides procedures for shutting down the CytoFLEX instrument.
Workflow:
Prepare the cleaning solution
This chapter contains information on:

Preparing the Cleaning Solution

Shutting Down the Instrument

Auto Shutdown [CytoFLEX LX Only]
Preparing the Cleaning Solution
Required materials
Materials to prepare:
• 12 x 75 mm sample loading tube
• FlowClean
• Deionized water
• Bleach
Set aside 2 mL of FlowClean in one sample tube and 3 mL of the deionized water in a separate sample
tube.
Shutting Down the Instrument
1
Run Daily Clean to clean the sample line. Refer to
in
CHAPTER 10, Cleaning
2
If necessary, empty all waste liquid from the waste container. Refer to
Container [CytoFLEX]
B49006AH
Ô
Clean the instrument
Daily Clean
Procedures.
in
CHAPTER 11, Replacement/Adjustment
CHAPTER 8
Daily Shutdown
Ô
Turn the instrument off
or
Daily Clean [With Plate Loader]
Emptying the 4 L Waste
Procedures.
8-1