VEE Measurement Program Operation
Measurement

Narrow Band TAA (Parametric)

Overview
This section describes the Narrow Band TAA measurement. Narrow Band TAA is
measured by the spectrum analyzer where low signal level measurement is possible.
Measurement
1. To open the Narrow Band TAA menu, click on the [Narrow Band TAA] button from
Procedure
Figure 3-44
Narrow Band TAA Menu
NOTE
Before you execute this measurement, you should set an appropriate input range or execute
the Auto Configuration to set the input range automatically.
2. Set the parameters, click on each button to enter a new value.
92
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
the Parametric frame of the Measure menu.
Channel Bit Rate
Write Current
Sense Current
Data Pattern
Average
Track #
Specifies the speed at which data bits are
written on the media per second [bps].
Specifies the current applied to the write head.
Specifies the current applied to the read head.
Specifies the data pattern to be written on the
test track. To set the data pattern, click on the
[Configuration] > [Setup] > [Pattern Config]
buttons to open the data pattern menu.
Specifies the number of revolutions to get the
average results.
Specifies the track number to be tested.
Chapter 3