VEE Measurement Program Operation
Setup Menu
4. Click on [Calibration] to start the amplitude calibration.
Figure 3-16
5. Replace the cable and adapter on E4402B again as advised on the display, and then
NOTE
It is recommended to do the calibration in case of the followings:
60
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Click on [OK] to finish the amplitude calibration.
a. Temperature has changed more than 5 deg. C after the previous calibration.
b. 24 hours or more has past after the previous calibration.
Chapter 3