Cassette and Z-Height Adjustment
Z-height Adjustment for E5029C/D Cassettes
Figure C-4
Zero Gauge Cassette
Figure C-5
Attaching Zero Gauge Cassette (Down Face)
NOTE
When the FIC length of cassette is adjustable such as E5029C Option C11, the letters "s"
(short length), "m" (medium length) and "l" (long length) is added as subscript to the
cassette's part number (e.g, "E5010-61145s", "E5010-61145m" or "E5010-61145l"). Refer
to Figure 3-3, "Cassette Parameter Setup Menu."
6. Rotate the micrometer knob upwards until the needle pin touches the contact block of
7. Check the LED of the Z-height tool. When the needle pin touches the contact block the
8. Push the "ZERO/ABS" button to set the zero height value to zero.
9. Remove the zero gauge cassette.
10. Rotate the micrometer knob upwards and set the micrometer reading to the required
394
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
the zero gauge cassette as shown in Figure C-5.
LED will turn on. At this stage, stop rotating the micrometer knob.