Hard disk read/write test system
Hide thumbs Also See for E5022A:
Table of Contents

Advertisement

Sense Curr. Step
Sense Curr. NOP
Off-track Start
Off-track Stop
Off-track Step
Off-track NOP
4.
Click on [Excite Config.] button to set the write excitation parameters of the SHNR
measurement.
Figure 3-97
Excite Configuration Menu
Excite Write Pattern
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
VEE Measurement Program Operation
Specifies the step by which sense current is
incremented while measuring SHNR. When
you set this parameter, Sense Curr [NOP] will
be computed and updated automatically.
Specifies the number of sense current points to
measure relative to the change in sense
current. When you set this parameter, [Sense
Curr. Step] will be computed and updated
automatically.
Specifies the starting point of the off-track
position in measuring SHNR.
Specifies the stopping point of the off-track
position in measuring SHNR.
Specifies the step by which the head is moved
in measuring SHNR. When you set this
parameter [Off-track NOP] will be computed
and updated automatically.
Specifies the number of measurement points.
When you set this parameter [Off-track Step]
will be computed and updated automatically.
Specifies the data pattern for write excitation.