Agilent Technologies E5022A Operation Manual page 394

Hard disk read/write test system
Hide thumbs Also See for E5022A:
Table of Contents

Advertisement

Figure C-2
Needle Pin Replacement
3. To replace the needle pin in Figure C-2, use a 0.71-mm hex screwdriver to loosen the
screw that locks the needle pin. Then select the pin hole to be used for measurement as
shown in Figure C-3.
Figure C-3
Needle Pin Position
4. Rotate the micrometer knob downwards to keep needle pin away.
5. Attach the zero gauge cassette (shown in Figure C-4) on the Z-height adjustment tool
completely. The user can use either sides (i.e, down face or up face) of the tool as long
as the locking pin of the cassette is compatible with the docking hole of the Z-height
tool as shown in Figure C-5.
Appendix C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Cassette and Z-Height Adjustment
Z-height Adjustment for E5029C/D Cassettes
393