Agilent Technologies E5022A Operation Manual page 7

Hard disk read/write test system
Hide thumbs Also See for E5022A:
Table of Contents

Advertisement

When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Contact your local Agilent Technologies sales representative or authorized service
company for repair of the instrument. If you continue to operate without
repairing the instrument, there is a potential fire or shock hazard for the operator.
— Instrument operates abnormally.
— Instrument emits abnormal noise, smell, smoke or sparklike light during
operation.
— Instrument generates high temperature or electrical shock during operation.
— Power cable, plug or receptacle on instrument is damaged.
— Foreign substance or liquid has fallen into the instrument.
Warning
Do not stick your fingers near the HLM- Spindle stage while it is moving as it may
cause injury to the operating personnel.
Do not stick your fingers in shroud while the spindle is running.
The furnished shroud does not provide protection from glass shards which can cause
death or serious injury. To prevent injury to the operating personnel use a safety cover
when you use glass media. Contact Agilent Technologies when you design a safety
cover.
Caution
Strong magnetic fields exist in the X-Y stage drive motors. Keep magnetic devices
away from the motor to avoid damaging it.
When applications such as VEE and Visual Basic is aborted abnormally or when the
reset button on the spinstand turns to red, you should remove the cassette first before
restarting. The cassette and spindle could bump into the shroud as the head moves to
home position during restarting.
Safety Symbol
General definitions of safety symbols used on the instrument or in manuals are listed
below.
Instruction Manual symbol: the product is marked with this symbol when it is necessary for
6