Cassette and Z-Height Adjustment
QM Stopper Replacement Procedure
2. To set the type of stopper to be used, first disassemble the HGA Mount block assembly
from the cassette as shown in Figure C-13. Use a 1.5-mm hex screwdriver to remove 2
hexagonal screws from the block assembly.
Figure C-14
Removing the stopper from the block assembly
3. Loosen the lock pin screw using a 0.71-mm (0.028") hex screwdriver.
4. Loosen the stopper screw using a 0.71-mm (0.028") hex screwdriver.
5. Remove the lock pin using a long nose pliers.
6. Attach the type of stopper to be used either QM/QS or Q30 type.
7. Insert the lock pin. The lock pin is slightly protruded outside the assembly.
8. Slowly tighten up the lock pin screw and stopper screw using a 0.71-mm (0.028") hex
screwdriver.
404