Agilent Technologies E5022A Operation Manual page 367

Hard disk read/write test system
Hide thumbs Also See for E5022A:
Table of Contents

Advertisement

Manual Changes
Service Release for Revision A.03.81
Problem Description:
In case of using E5029H Opt.015, when the measurement interval is more than about 1
minute, next first measurement result is not good. After one bad measurement result, the
result becomes good if the interval is less than about 1 minute.
Cause: At the revision 3.81 and below, the SR1750 was set at IDLE mode at unload
position. This causes the problem. At this service release, the SR1750 was set at SLEEP
mode at unload position.
366
Appendix A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com