Cassette and Z-Height Adjustment
Z-height Adjustment for E5029C/D Cassettes
Figure C-6
Adjusting the height of the cassette
12. Loosen the flat screw located on top of the cassette as shown in Figure C-6 until the
Z-height is almost at its highest position. Use a slotted screw driver.
13. Attach the cassette to the Z-height adjustment tool as shown in Figure C-7.
Figure C-7
Attaching the Cassette
14. Slowly tighten the flat screw until the LED turns on.
396