Agilent Technologies E5022A Operation Manual page 400

Hard disk read/write test system
Hide thumbs Also See for E5022A:
Table of Contents

Advertisement

2. Remove the battery cover (black button) by turning it counter-clockwise using slotted
screwdriver or coin.
3. Remove the battery and replace it with a 1.55-V button type battery (SR44 or LR44).
This battery is commercially available. Typically this battery lasts for one year.
4. Replace the battery cover by turning it clockwise using a slotted screwdriver or coin.
5. Turn the Z-height tool down face and remove 6 hexagonal flat screws from the bottom
cover of the Z-height tool using 2-mm hex screwdriver as shown in Figure C-10.
Figure C-10
Removing the bottom cover
6. Unscrew 3 hexagonal round screws from the pc board using a 2-mm hex screwdriver as
shown in Figure C-11.
Appendix C
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Battery Replacement Procedure for Micrometer and Z-height Tool
Cassette and Z-Height Adjustment
399