[Optimize] Menu
Measurement
Click on the [Optimize] button to open the optimization menu.
Procedure
Optimization enables the channel IC to operate at optimum performance. Optimization
is necessary whenever you change the user data rate, endec parameters, etc. prior to
measurement. Optimization enables you to operate the Equalizer and filter settings at
maximum performance. It also allows you to evaluate the write precompensation.
Exec. Optimize
5. Set the parameters, click on each button to enter a new value.
Data Rate
BER Pattern #
Track #
Optimize Type
Optimize Sequence
Sequence Setup
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
VEE Measurement Program Operation
Optimizes the selected parameters in the
[Sequence Setup] menu.
Specifies the number of data bits to be
recorded per second.
Specifies the pattern number to be used to
write the data pattern. The data pattern of the
selected pattern number must be specified in
the [Pattern Config.] menu in advance.
Specifies the track number to be tested.
Selects the sequence type. This selects if it
executing a erase and a write before the
optimization. When the "Optimize Only" is
selected, the optimization of write
precompensation must be turned off.
Specifies the type in the optimization. Allows
the user to select either 'Standard" or
"Custom" optimization.
A configuration menu will pop up when you
Measurement
167