system error "hpe5022_ERROR_NSUP_CONF" will be displayed, when you set the
[Write Mode] to "Continuous".
Figure 3-35
Write Mode
Write Pattern
Write Mode
5. Click on the [Measure] button to start measurement. Results are displayed after
measurement.
Measurement
The menu shows the measurement results of TAA, TAA+ and TAA- from the first
Display
revolution up to the # of Rev. The mean, minimum, maximum and standard deviation
of the TAA are all displayed on the upper portion of the graph.
6. Click the [Return] button to return to the Stability menu.
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
VEE Measurement Program Operation
Specifies the write excitation pattern for
stability measurement.
Specifies the write data polarity when writing
the data. When 'Fixed' is selected the write
current direction of each write operation is the
always the same for all segments. When
'Continuous' is selected the write current
direction at the beginning of write operation is
the same as the last current direction of the
previous write. Refer to the programming
manual for details.