Expiratory Trigger (Exp.trigger); Rise Time - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

Property
Setting resolution
Available in modes

4.2.10 Expiratory Trigger (Exp.Trigger)

The Expiratory Trigger setting defines the moment when the ventilator will cycle from
the inspiratory to the expiratory phase.
Property
Value range
Default value
Setting resolution
Available in modes
4.2.11

Rise Time

The Rise Time setting controls the speed of the pressure increase from start of inspira-
tion to set inspiratory pressure.
A low setting will give a faster pressure increase and therefore a longer plateau at the set
inspiratory pressure. A high setting will give a slow increase and therefore a shorter
plateau.
Property
Value range
Default value
Setting resolution
Available in modes
Doc. 007232 Y-2
Description
1
• PCV+A
• PSV
• S
• S/T
Description
Auto, 1 to 9, corresponding to 10% to 90% of peak flow
Auto
1
• PSV
• S
• S/T
Description
Auto, 1 to 9
Auto
1
• PCV+A
• PSV
• S
• S/T
• T
Treatment Functions and Settings 57