5.4.2

Disconnect the Heated Patient Circuit from the Ventilator

1
Disconnect from the ventilator:
Press the latches and pull the circuit off from the ventilator. Do not pull by the tube.
2 Disconnect from the patient interface:
Hold by the cuff and pull the circuit off from the patient interface. Do not pull by the tube.
5.5

Performing Start-up Checks

This procedure may be performed as a functional pre-use check at start-up.
Procedure
Prerequisites
The Vivo 3 shall be connected to the power supply.
1
Start the treatment.
• If Confirm Start/Stop is Off, press On/Off button shortly.
• If Confirm Start/Stop is On, press and hold the On/Off button until the Starting
For information about setting the Confirm Start/Stop options, see 3.5.5.1 Device Settings, page
28 and 6.2 Start the Treatment, page 40.
.
2 When starting treatment, an automatic alarm signal test is performed. Check that the
test is performed successfully, this is indicated by:
• A short beep indicating functional audio signaling.
• The alarm LED first lights yellow, then red, indicating functional visual signaling.
• The audio pause LED lights yellow.
• In about a second, both LEDs are turned off.
If the test fails, do not use the Vivo 3. Contact your supplier of the Vivo 3 for a technical
check.
3 Disconnect the power cord for more than 5 seconds.
Check that the device switches to the internal battery and that the information message
Switched to Internal Battery is shown on the screen.
If not, contact your supplier of the Vivo 3.
4 Reconnect the power cord.
Check that the Vivo 3 switches to the mains supply.
If not, contact your supplier of the Vivo 3.
5.6

Performing a Pre-Use Test

The pre-use test is used for detecting the characteristics of the patient circuit that is con-
nected to the ventilator. The resistance and compliance of the patient circuit are measured
and calculated. This will be used to compensate for pressure drop in the patient circuit and
the compliance of the patient circuit.
The patient shall not be connected during the pre-use test.
38 Prepare the Vivo 3 for Use
User Manual