Change Of Patients; Alarm Tests; Mandatory Test - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

8.3

Change of Patients

If assigning the Vivo 3 to a new patient, first clean and prepare the Vivo 3 as described
in this section.
If a humidifier is used, it is classified as single patient use.
NOTE
If the Vivo 3 is used in a clinic, with frequent change of patients, a low resistance bac-
teria filter is located between the air outlet and the patient circuit to prevent patient
cross contamination.
1
Clean the outside of the Vivo 3 according to 8.1.1 Clean the Main Unit Externally,
page 109.
2 If a low resistance bacteria filter has been used between the patient air outlet and the
patient circuit: Replace the filter.
If a low resistance bacteria filter not has been used, the airways of the Vivo 3 may
need disinfection to prevent patient cross contamination. See 8.1.2 Air Pathway Dis-
infection, page 110.
3 Replace the patient air inlet filters according to 8.2 Clean and Replace Patient Air Inlet
Filters, page 110.
4 use a new patient circuit for the new patient.
8.4

Alarm Tests

The alarm tests should be performed when changing patient, if the ventilator's function
needs to be checked for any other reason, or at least every 12 months.
8.4.1

Mandatory Test

8.4.1.1 Low Pressure Alarm Test
This test ensures safe use of the Vivo 3.
1
Connect the Vivo 3 to a test lung.
2 Take notes of the current settings, if they are to be used also after the test.
3 Adjust the setting as follows:
Setting
Ventilation mode
IPAP
EPAP
112 Cleaning and Maintenance
Vivo 3 Clinician's Manual
Value
PSV
15 cmH
O
2
5 cmH
O
2
Doc. 007232 Y-2