Cleaning And Maintenance - Warning And Precautions - Breas Vivo 3 Clinician Manual

Hide thumbs Also See for Vivo 3:
Table of Contents

Advertisement

Risk of Suffocation
Installation of a water trap may be required if the condensation is extensive in the
patient circuit when using a heated humidifier.
The water trap prevents condensed water in the patient circuit from reaching the
patient airways and causing personal injury.
Risk of Suffocation
Do not use the attachable humidifier during mobile use.
Due to movements, water spillage from the humidifier or condensed water may flow
to the patient and cause suffocation.
WARNING!
Risk of Electric Chock
Do not use the attachable humidifier during mobile use.
Internal water spillage may cause electric chocks and may damage the device.
Risk of electric shock
If using the protective cover or the carry bag, first remove the attachable water cham-
ber. Water spillage may cause electric shocks.
Risk of Electric Shock
Electrostatic discharge or contact with leakage current may cause an electric shock.
The attachable humidifier must be detached before filling. Do not fill above the Maxi-
mum Water Level indication on the water chamber.
WARNING!
Risk of Burns
After using the ventilator, wait one minute before you open the water chamber since
it can get hot under certain conditions (for example if the humidifier runs out of
water).
CAUTION!
The use of external humidifier may require readjustment of the low-pressure alarm.
2.8
Cleaning and Maintenance — Warning and Precautions
This manual contains instructions for cleaning and maintenance that can be carried out
by the care provider or users with physical ability and working knowledge of the system.
Doc. 007232 Y-2
Safety Information 27
Vivo 3 Clinician's Manual

Advertisement