Disposable Filter (white)
Replace the white filter at least every 4
tion or pollen-rich environments.
CAUTION!
Do not wash or reuse the disposable filter.
7.3

Change of Patients

If the ventilator is used in a clinic by several patients, a low resistance bacterial filter may be
used between the air outlet and the patient tube to prevent patient cross-contamination.
1 Follow the instructions in 7.1 Cleaning the Vivo 65, page 124, steps 1 to 5.
2 Replace the patient filters according to 7.2 Cleaning and Replacing the Patient Air Filters,
page 126.
3 If a low resistance bacterial filter is used, it shall be replaced..
4 Use a new patient circuit when the ventilator is used by a new patient.
7.4

Regular Maintenance

Regular maintenance inspections and checks shall be carried out at least every 24 months,
according to the ventilator Service Manual.
WARNING!
Do not use the device and contact your responsible care provider for an inspection of
the device in the event of:
7.5

Service and Repair

The service and repair of the ventilator must only be carried out by authorised service per-
sonnel in accordance with Breas service instructions. Service inspections must always be car-
ried out following any repairs to the device.
Doc. 006150 M-1
Unexpected patient symptoms during treatment.
Unexplainable or sudden pressure, performance or sound changes during
operation.
Suspected damage to the device, including the occurrence of Internal Functional
Failure alarms.
Suspected damage to the click-in battery, including evidence of battery cell leakage.
th
week, or more frequently when used in high pollu-
Cleaning and Maintenance