Cleaning and Maintenance
This chapter contains instructions for cleaning and maintenance actions that can be carried
out by the care provider or by users with physical ability and working knowledge of the
system.
WARNING!
Risk of Personal Injury
•
•
The patient-connected parts and the filters must be cleaned and replaced regularly to ensure
correct function of the Vivo 3. All replaced parts must be disposed of in accordance with
local environmental regulations regarding the disposal of used equipment and waste.
8.1
Cleaning the Vivo 3
WARNING!
Risk of Electric Shock
Disconnect the power supply before cleaning theVivo 3 according to the instructions in
this manual.
Do not soak the Vivo 3 or immerse it into any fluids.
8.1.1
Clean the Main Unit Externally
Equipment
•
A lint free cloth.
•
A mild soap solution or Ethanol 70%.
1
Turn off the Vivo 3 and disconnect the power supply.
2 Remove the patient circuit.
3 If any cable connected accessories (like the SpO
disconnect them.
4 Clean the outside of the Vivo 3 using a lint free cloth moistened with a mild soap
solution and / or ethanol 70%.
5 When the equipment is clean and dry, reconnect the patient circuit and any accessories
that was disconnected during the cleaning.
Doc. 007231 F-3
Repairs, upgrades and modifications must be carried out by technicians authorized
by Breas Medical AB only and in accordance with instructions from
Breas Medical AB
The Vivo 3 must not be opened, repaired or modified by unauthorized personnel.
If subjected to unauthorized operations, Breas Medical AB is no longer responsible
for the performance and safety of the device and all warranties will become invalid.