8

Cleaning and Maintenance

This chapter contains instructions for cleaning and maintenance actions that can be car-
ried out by the care provider or by users with physical ability and working knowledge of
the system.
WARNING!
Risk of Personal Injury
• Repairs, upgrades and modifications must be carried out by technicians authorized
by Breas Medical only and in accordance with instructions from Breas Medical
• The Vivo 3 must not be opened, repaired or modified by unauthorized personnel.
If subjected to unauthorized operations, Breas Medical is no longer responsible
for the performance and safety of the device and all warranties will become invalid.
The patient-connected parts and the filters must be cleaned and replaced regularly to
ensure correct function of the Vivo 3. All replaced parts must be disposed of in accord-
ance with local environmental regulations regarding the disposal of used equipment and
waste.
8.1

Cleaning the Vivo 3

WARNING!
Risk of Electric Shock
Disconnect the power supply before cleaning theVivo 3 according to the instructions
in this manual.
Do not soak the Vivo 3 or immerse it into any fluids.
8.1.1

Clean the Main Unit Externally

Equipment
• A lint free cloth.
• A mild soap solution or Ethanol 70%.
1
Turn off the Vivo 3 and disconnect the power supply.
2 Remove the patient circuit.
3 If any cable connected accessories (like the SpO
used, disconnect them.
4 Clean the outside of the Vivo 3 using a lint free cloth moistened with a mild soap
solution and / or ethanol 70%.
5 When the equipment is clean and dry, reconnect the patient circuit and any accesso-
ries that was disconnected during the cleaning.
Doc. 007232 Y-2
sensor or the accessory box) are
2
Cleaning and Maintenance 109
Vivo 3 Clinician's Manual