6.3.2

Low Pressure Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
Setting display
92
Alarms
User manual
Description
Low Pressure
High
A Low Pressure alarm will be given when the Vivo 65 pressure
fails to reach the low pressure alarm limit for 15 seconds.
In MPV breath mode the alarm will be given when the pressure
fails to reach the limit during inspiration.
Disconnection of patient circuit.
Mismatch between pressure setting and alarm setting.
Leakage from the mask or other components of the patient
circuit.
The pressure rises above the alarm limit.
The Vivo 65 will continue treatment according to the current
settings.
1 cmH
O to 60 cmH
2
Note that the Low pressure alarm cannot be set higher than
the value set for the High pressure alarm.
Below 10 cmH
O: 0.5 cmH
2
Above 10 cmH
O: 1.0 cmH
2
The Low Pressure alarm setting is displayed by a red line in the
pressure bar graph.
O
2
O
2
O
2
Doc. 006150 M-1