Download Print this page

External Gain - Keysight Technologies C Series User Manual

Fieldfox analyzers

Advertisement

SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
SA Mode Settings
How to Set Y-axis Units
The Unit setting appears for the reference line, marker readouts, and trigger
level. All Unit choices are available in both Log and Linear Scale Types.
The UNIT setting survives Mode Preset, but NOT Preset.
— Press Scale / Amptd, then More, then
— Then choose from the following:
Antenna correction units are available ONLY by loading or editing an Antenna
file that contains the desired units setting. Learn more in
page

External Gain

When using an external amplifier or attenuator, the SA mode trace amplitude
values can be offset to compensate for the effect of the external device. This
effectively moves the reference plane of the SA measurement port out to just
beyond the external device. For example, when using an external preamp with
gain of +10 dB, enter this value for External Gain, and the data trace across the
displayed frequency span will be adjusted down by 10 dB.
When RF Atten is set to Auto, you may see a change in the RF Attenuation
value. This is an attempt to measure the signal at top of screen (the
acquisition) without overloading the SA first mixer.
How to Set External Gain
1. Press Scale/Amptd.
2. Then
3. Then
4. Enter a value using the numeric keypad, the
160
PSD Function
Enables/disables the Power Spectral Density Function (PSD
Function) which is a way to normalize the SA measurement to a 1 Hz
Bandwidth. (Default == OFF). Learn more, see
(PSD) Function (SA Mode Only)" on page
—dBm (default setting), dBmV (dB milli Volts), dBuV (dB microVolts), W
(watts), V (volts), A (amps), dBmA (dB milliAmps), dBuA (dB
microAmps).
826.
More
External Gain
knob (positive for gain; negative for loss). Values less than 5 dB must be
typed using the numeric keypad. Then press Enter
ExtGain xx dB
is shown at the top of the screen.
"Power Spectral Density
216.
Unit
"Manage Files" on
arrows, or the rotary
Keysight N9915-90020 User's Guide

Advertisement

loading

This manual is also suitable for: