SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
SA Mode Settings
— Noise Marker Averaging - Learn more in
— Video BW filtering (for non-Zero Span measurements) - Learn more in
There are three types of mathematical averaging that can be performed – Log,
Linear, or Voltage (see
of these types and it is used for all of the above averaging processes.
How to set Average Type
1. Press Meas Setup 4 >
2. Then choose from the following:

Average Count

The Average Count setting is used mainly with the Average Trace State
described in
State, the Average Count setting determines the number of sweeps to average.
The higher the average count, the greater the amount of noise reduction.
When Trace (display) State is set to Average, MaxHold, or MinHold, the
average counter is shown in the left edge of the screen below the Average
Type.
For all three of these Trace States, when Sweep 3
press
to Hold.
How to set Average Count
1. Press Meas Setup 4 >
Keysight N9915-90020 User's Guide
"Video Bandwidth (Video BW)" on page
"How to set Average Type" on page
Auto
- The FieldFox chooses the most appropriate type of averaging
for the current settings. When Detection Method is set to Average, a
Noise Marker is present, or a Channel measurement is active, then
Power Average is ALWAYS selected. Otherwise, Log Average is
selected.
Log
Averaging – Best for displaying Trace Averaging.
shown on the left side of the FieldFox screen when selected.
Power
(Linear) Averaging – Best for measuring true power levels.
Used in Detection Average and Noise Marker Average.
Mathematically, trace noise is 2.5 dB higher than when using Log
Average. PAvg is shown on the left side of the FieldFox screen when
selected.
Voltage
Averaging – Converts the amplitudes to a magnitude unit
before averaging (e.g., Volts, Amps, microVolts, milliVolts, etc.)
#VAvg
annotation is displayed.
"Trace Display States (SA Mode)" on page
Restart
to reset the sweep count to 1, perform <n> sweeps, then return
"Noise Marker" on page
188.
Average Type [current setting]
Average Count
215.
205). Select ONE
#LgAv
is
203. In this Trace
Continuous
is set to OFF,
205