On Fail
—
Off
—
Warning On Off
points that fail are highlighted.
—
Limit Type
Then Fixed or Relative.
—
Offset
Shifts the limit line up or down. Enter an offset value using the rotary knob, the
arrows, or by entering a value with the number keys.
— Enter a positive offset to shift the limit line UP, usually used with UPPER limits.
— Enter a negative offset to shift the limit line DOWN, usually used with LOWER
limits.
—
Margin
Shifts the limit line up or down like an offset, but only virtually. The limit line
does NOT appear to move. Enter a Margin value using the rotary knob, the
or by entering a value with the number keys.
Enter a positive margin to virtually shift the limit line UP.
Enter a negative margin to virtually shift the limit line DOWN.
—
LogX Interp Auto Off
(default == Auto). Learn more, refer
233–Mixed Analyzers)FUTURE"
modes Only)" on page
How to Save and Recall Limits
1. Press Save/Recall 9 then
used for the Save or Recall. Learn more in
2. Press Limit 8.
3. Then choose from the following:
—
Save Limits
After limit lines have been defined, this saves the line definition to a file on
the specified device (e.g., *.xml).
—
Recall Limits
Recall can recall both *.xml or *.csv file types. But, Save can only save files
as *.xml.
754
Data Analysis FeaturesDELETE_IF
All about Limit Lines
Audible beep sounds when a sweep fails.
No audible beep sounds.
Shows Pass or Fail indication on the display with each sweep. Data
– Enables/disables the logarithmic X-axis interpolation trace
and to
217.
Device
to set the Device (Internal, SD card, or USB flash drive) to be
Recalls a limit Line definition from the specified device (e.g., *.csv).
Chapter 9, "SA (Spectrum Analyzer) Mode (Option
"Logarithmic X-axis Interpolation (EMI and SA
"Set File Type and Select Device" on page
Keysight N9915-90020 User's Guide
arrows,
823.