Download Print this page

Keysight Technologies C Series User Manual page 817

Fieldfox analyzers

Advertisement

System SettingsFUTURE
Service Diagnostics
Erase User Data
IMPORTANT!
The process in this section erases all User Data from the FieldFox. Before
proceeding, Keysight strongly recommends that you back up all of your
User Data to an external storage device. refer to
page
All data that you have written to the FieldFox can be erased with the press of a
button. You would do this if your FieldFox is located in a secure environment
and must be routinely erased of sensitive data.
All data is removed from the FieldFox internal "UserData" partition. This
includes all State files and Trace + State files (*.sta), all Data files (*.csv, *.snp),
all Picture files (*.png), all Cal Kit files, and all Cable files.
External media that is inserted (mini-SD card or USB flash drive) is NOT
erased.
The "Lost Clusters" folder and files are NOT erased because they are part of
the file system housekeeping.
How to Erase User Data
1. Press System 7 >
2. Then
— Then the following occurs:
Keysight N9915-90020 User's Guide
807.
Service Diagnostics
Confirm Erase
— All data files and folders are deleted from the "UserData" partition.
— The FieldFox is rebooted, which manages the newly-freed data and
re-writes the factory cal kits and cable files.
"How to backup files" on
>
Advanced
>
Erase User Data
813