SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
Spectrum Emission Mask (SEM) — (SA Mode Only)
4. Setting the Limit Lines
The reference power level is displayed in green. The reference power level
type can be selected through Measure 1 >
power settings defined through Meas Setup >
Example: If the reference power level reads -15 dBm, and
to -10 dB, the level is -25 dBm. Since this limit line type is relative, it
moves with the reference power reading.
248
Two Limit traces are available, an Absolute Limit and a Relative Limit. Both
are defined and applied only over the Offset measurement regions. There
is no Limit definition or application for the Reference Channel itself. The
final Limit Trace displayed is a concatenation of all the individual Offset
Limit segments, just as the final measurement trace is. Success or Failure
is determined by Fail Mask per offset, to be based on either Relative or
Absolute limits, Both (Abs AND Rel), or Either (Abs OR Rel).
Press Limit 8 then
Offset
- selects the offset region to be edited.
The Limits per offset are determined by Start/Stop Amplitude level
endpoints, which correspond exactly to the user specified
Start/Stop frequency locations in the Offsets sub-menu. So each
offset segment is a LINE defined by its two endpoints. In the Limit
"Auto" means to match the Start limit value, hence creating a FLAT
(horizontal) Limit Line segment. But the Stop value can be manually
set as well, creating sloped Limit segments.
Abs Start
- Sets the start of an absolute limit line segment.
Abs Stop
- Sets the stop of an absolute limit line segment.
Rel Start
- Sets the amplitude limit at the start frequency of the
offset, relative to the reference power level.
Rel Stop Auto
- Sets the amplitude limit at the stop frequency of the
offset, relative to the reference power level.
Fail Mask
- Sets the Fail Mask offset parameters for the
measurement. Then choose:
Absolute
Relative
Abs OR Rel
generates a fail message.
Abs AND Rel
failure to generate a fail message.
Meas
Ref
- The absolute mask is tested for all offsets.
- The relative mask is tested for all offsets.
- Either Absolute or Relative data failure
- Requires a relative and an absolute data
Keysight N9915-90020 User's Guide
Type, and the reference
Channel.
Rel Start
is set