External Temperature Measurement; Dimensions And Weight - Keysight Technologies BT2152B Operating And Service Manual

Self-discharge analyzer
Table of Contents

Advertisement

External Temperature Measurement

The external temperature sensor input on the rear panel requires an external 10 kΩ thermistor.
Measurement Input
Thermistor Requirements:
Temperature measurement range:
Temperature measurement uncertainty:
Maximum total lead resistance allowed:

Dimensions and Weight

Do not block the air intake at the front and side, or the exhaust at the rear of the instrument.
The front air inlet requires a minimum of 6 inches (15 cm) clearance. Inlet air temperature must be
l
within the temperature range of 20°C to 30°C.
The rear air outlet requires a minimum of 6 inches (15 cm) clearance.
l
The small air outlet on the left side of the unit must also be unrestricted, with a minimum clearance of
l
1.25 inches (30 mm) or more. Clearance to the instrument support rail could be as small as 10 mm if
rack mounted.
The instrument should not be installed in a pressured rack or enclosure.
l
Weight
10.2 kg ((22.5 lbs.)
Keysight BT2152A/B Operating and Service Guide
Characteristic
Negative Temperature coefficient (NTC)
10 kΩ Nominal Resistance at 25° C
Programmable Beta value (4073 default)
5° C to 100° C
± 1.5° C
10 Ω