CAT (Cable and Antenna Test) Mode
CAT Mode Settings
Scale annotation on the FieldFox screen
Reference Line = red arrow
Ref Level = -40 dB
Ref Position = 1
Scale = 2 dB per division

Averaging

Trace Averaging helps to smooth a trace to reduce the effects of random noise
on a measurement. The FieldFox computes each data point based on the
average of the same data point over several consecutive sweeps.
Average Count determines the number of sweeps to average; the higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as Avg N. This
shows the number of previous sweeps that have been averaged together to
form the current trace. When the counter reaches the specified count, then a
'running average' of the last N sweeps is displayed. Average Count = 1 means
there is NO averaging.
54
Autoscale
1.
Automatically adjusts the Y-axis to comfortably fit the
Minimum and Maximum amplitude of the trace on the screen.
2. Set Scale, acquisition, and Reference Position:
Scale
Manually enter a scale per division to view specific
areas of the trace
Ref Level
Enter a negative value by pressing Run/Hold (+/-) either
before or after typing a value
Ref Pos
Manually set the position of the reference line.
Values must be between 0 (TOP line) and 10 (BOTTOM
line)
3. Set Top and Bottom graticule values. The scale per division is
calculated.
Top
to set the value of the Top graticule.
Bottom
to set the value of the Bottom graticule.
— Enter a negative value by pressing Run/Hold (+/-) either
before or after typing a value.
Manually set the value of the reference line.
Keysight N9915-90020 User's Guide