Keysight Technologies C Series User Manual page 371

Fieldfox analyzers

Advertisement

Over–the–Air Measurements (OTA) – LTE FDD Option 370, LTE TDD Option 371, 5G TF
Option 377, 5G NR / 5G NR EVM Conducted Option 378
How to Set Up Over–the–Air (OTA) Mode Settings
Keysight N9915-90020 User's Guide
— Press
Auto-Range
the optimal RF attenuation and preamp settings for the
current input signal.
Executing an auto-range does NOT change any of the display
settings (i.e., Ref Level, Scale or Ref Pos).
See also,
"About the Auto-Range Feature" on page
— Press
External Gain
b. Optional (When Display Type = Bar Chart): Modify scale and specify
color bar limit values—Modify the bar chart display color coding
scheme for the power channels (bars):
Press
Trace
>
More
Then one of the following:
CC
to set the CC (Component Carrier) associated with each
bar.
PCI
to set the PCI (Physical Cell Identifier) associated with
each bar.
Level
to set the level of the value associated with each bar.
When level is selected, you can set the channels' red and blue
power level limits with the Scale/Amptd softkeys:
Scale/Amptd
Press
Then one of the following:
Red Limit
to set the power level and color scale for high
power channels.
Blue Limit
to set the power level and color scale for low
power channels.
— Others: Scale,
More
opens a menu of softkeys that are the same as those
that are defined when the Display Type = Table.
— Others (optional):
c. Optional (When Display Type = Spectrum): Modify RF attenuator,
Preamp, and other settings:
Press
Scale/Amptd
Then press the following:
Ref Pos
Manually set the position of the reference line. Values
must be between 0 (TOP line) and 10 (BOTTOM line).
Scale
Manually enter a scale per division to view specific
areas of the trace.
Executes an auto-range, which determines
>
Bar Color Coding
Ref
Level, etcetera as desired.
Autoscale
412.
371