Pulse Measurements Mode – Option 330
Pulse Measurement Settings
— Then
— Then choose from the following:
Then choose from the following:
— Target, Peak, and Minimum. Learn more about these standard features
Rise and Fall time in Marker Search uses different criteria than in Auto
Analysis and will therefore result in different reported values.
Auto Analysis reports the following TEN pulse measurements:
For more information on peak power measurements refer to the
documentation for your power sensor. Learn more,
https://www.keysight.com/us/en/assets/7018-03076/application-notes/
5990-8708.pdf.
1 - Average Power
2 - Peak Power
3 - Peak minus Average
4 - Pulse Top
5 - Rise Time
606
Zone Search
ON
Available ONLY when the zoom window is enabled, the search is
limited to the zone defined by the zoom window.
OFF
The search is performed over the entire Tr 1 trace.
beginning in
"Searching with Markers" on page
Rise Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value BEFORE the Peak.
Fall Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value AFTER the Peak.
Auto Analysis
Average power level of all measured data
points
Highest reading of all measured data
points
Difference between the above two results
Power value based on the specified Pulse
Top setting.
The time it takes for a pulse to rise from
10% to 90% of its peak value.
Keysight N9915-90020 User's Guide