SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
SA Mode Settings
If FM, choose from:
If FM Stereo / RDS, choose from:
If PM, choose from:
For any of the following: AM DSB, AM SSB, FM, FM Stereo / RDS, or PM,
choose from:
Refer to
page
— Then, press
Keysight N9915-90020 User's Guide
FM Top Y
– Determines the Y axis frequency (1 kHz to 1 GHz) of the FM
demodulation waveform is displayed (Default = 100 kHz).
FM Top Y
– Determines the Y axis frequency (1 kHz to 1 GHz) of the FM
demodulation waveform is displayed (Default = 100 kHz).
PM Top Y
– Determines the Y axis phase (0 to 100 radians) of the PM
demodulation waveform is displayed (Default = 5).
Figure 9-7 on page
183, and
Figure 9-10 on page
Demod SwpT
– Sets the X-axis to display the demodulation sweep time for
the demodulated waveform (minimum = 50 μs; maximum = 50 ms).
Peak OFF ON
– This adds peak positive and peak negative metrics to the
main metrics measurement. For example, in FM peak ON will include peak
positive (Peak +) and negative (Peak -) frequency deviation metrics to
appear. Peak ON will also enable peak traces on the demodulation
waveform and audio frequency spectrum (Default = OFF).
Demod Type
– Sets the demodulation type.
Choose from AM,
AM
More
– opens a menu of softkeys:
Count, and
Demod View Time
Average Type
– Selects the way the demodulated signal values are
averaged. Choices are Power (#PAvg annotation is displayed), Log (#LgAv
annotation is displayed), Voltage (#VAvg annotation is displayed). Auto is
the default and when selected the FieldFox chooses the type of averaging
that is used based on the current settings.
Average Count
– Determines the amount of averaging used (1 to 10K;
Default = 10).
Demod Filters
– Refer to
Demod View Time Freq
displayed (Time (waveform) or Frequency (spectrum)); Default = Time).
Audio Capture Playback
Softkeys" on page
178.
182,
Figure 9-8 on page
184.
SSB, FM,
FM Stereo /
Freq.
"How to Set the Demod Filters" on page
– Toggles how the demodulated signals are
"Setting the Audio Capture Playback
– Refer to
183,
Figure 9-9 on
RDS, and PM.
Average
Type,
Average
180.