Download Print this page

Preamplifier Control (Opt 235); Check For Compressed Measurements - Keysight Technologies C Series User Manual

Fieldfox analyzers

Advertisement

SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)FUTURE
SA Mode Settings

Preamplifier Control (Opt 235)

When very low-level signals are analyzed, an internal preamplifier can be used
to boost the signal level by approximately 22 dB. The gain of the preamp is
NOT adjustable. The displayed signal level is automatically adjusted for the
increase in system gain.
How to control the Preamp
By default, the preamp is OFF.
— Press Scale/Amptd.
— Then

Check for Compressed Measurements

Compression occurs when too much power into an amplifier causes it to no
longer amplify in a linear manner. When too much power goes into the
FieldFox RF Input connector, the amplifiers in the SA receiver compress and
signal power will not be displayed accurately. This can occur even if
Range
might be compressed. Increase the RF Attenuation value to prevent the SA
receiver from being compressed.
How to Check for a Compressed Measurement
1. Using a marker at the signal peak, make note of the signal power level.
2. Increase the RF Attenuation level by 5 dB.
162
Man
RF Attenuation is set manually. The default Attenuation setting
is 10 dB. Enter a value between 0 to 30 dB in 5 dB steps using the
numeric keypad, the
Enter
#Atten xx dB
is shown at the top of the screen. (#) means manual
setting.
Preamp ON OFF
ON,
PA
is shown to the left of the graticules on the display.
is not displayed because devices other than the ADC, such as the mixer,
— If the signal level does NOT change, then NO compression exists.
This indicates that the signal is in the linear region of the FieldFox
receiver.
— If the signal level DOES increase, then the receiver was compressed.
Set the RF Attenuation value at the setting when further increases
no longer result in an increase in the displayed power level.
arrows, or the rotary knob. Then press
The underlined setting is the current setting. When
Keysight N9915-90020 User's Guide