How to move an IQ and SA mode marker between traces
1. With two or more traces present, press Marker and Enter to create a marker. The marker is
created on the current Marker Trace selection (default setting is 1).
2. Then
More
3. Then select the marker to move using
4. Then toggle
Marker Trace 1,2,3,4
— The selected marker is moved to the specified trace.
— When a marker is moved to a 'blank' trace (not visible) the marker moves to the same
X-axis position at either the Y-axis position of the last data, or to the ZERO position.
Marker Format
The Marker Format setting allows you to choose the format of the marker readout that appears in
the upper-right corner of the screen and in the marker table. The marker format can be
independent of the trace format, and can be set independently for each trace. All markers on a
trace have the same marker format.
How to choose marker format
— Press Marker
— Then
More
Marker Format
— Then
— For NA Mode measurements, choose from the following:
Selection
Default
R + jX
(3rd number is distance for Time Domain)
Z Magnitude
Phase
Real
More
Imaginary
Mag & Phase
dB Angle
Keysight N9915-90020 User's Guide
Data Analysis FeaturesDELETE_IF
All about Markers
Marker
1,2,3,4,5,6.
Same as displayed format
Complex impedance format
Impedance Mag.
Example Y-axis value
22.8Ω -j61.4Ω
287.9fF
66.08Ω
73.8°
0.2003
-0.6727
0.705, -73.7°
-2.1 dB, -138.2°
739