Replacing Gui Software Chip, Only For Black/White Gui - Planmeca ProMax Technical Manual

Hide thumbs Also See for ProMax:
Table of Contents

Advertisement

REPLACING THE GUI SOFTWARE

3.2 Replacing GUI software chip, only for black/white GUI

WARNING
Always turn the X-ray unit off before removing the software chip
from its socket.
Never turn the unit on, if the software chips are not in their sock-
ets.
NOTE
Antistatic precautions must be performed when handling the software chip. Touch any
grounded metal part of the unit before touching the software chip.
NOTE
The GUI software update by changing the software chip can be ONLY done for a black/
white GUI.
Detach the user interface according to the instructions given in section 2.2 "Support arm
lower cover" on page F-8.
Remove the software chip from its socket marked IC16 with the special tool.
IC 1
Figure 14
NOTE
If an IC1 chip is present, remove it too.
Carefully place the new software chip into the socket.
Attach the user interface to the patient support arm.
Technical Manual
IC 16
+
FLASH 3
FLASH 4
Chapter F - PATIENT SUPPORT ARM
FLASH 2
FLASH 1
GUI
Planmeca ProMax F-11