ADJUSTMENTS AND CALIBRATIONS
Figure 6

1.2 Preparations before adjustments

Switch the unit off. Remove the housing of the cephalostat head support, see section 4.1
"Removing the housings of the head support" on page H-64.
Remove the second primary collimator according to the instructions given in section 2
"REMOVING AND ATTACHING THE SECOND PRIMARY COLLIMATOR" on page H-55.
Remove the tube head cover and the C-arm inner cover, refer to section 5 "REMOVING THE
COVERS" on page E-77.
If the cephalostat is equipped with movable sensor head, detach the sensor head and
remove the sensor head connector covers according to the instructions given in section 3.1
"Cephalostat with movable sensor head" on page H-57.
If the cephalostat is equipped with fixed sensor head, see section 3.2 "Cephalostat with
fixed sensor head - removing and attaching the sensor head" on page H-61.
Technical manual
3D Flat Field calibration tool (part number 10017348).
Chapter H - CEPHALOSTAT
Planmeca ProMax H-3